- Home
- Tuotteet
- Juomateollisuus
- Steinfurth
- Pakkausanalyysi
- Steinfurth Closure Tester (CT V6)
Steinfurth Closure Tester (CT V6)
- General Details
- Product Details
- Advantages
- Technical Data
- Documents
Resistance against interior pressure should be tested frequently, but at least when a new closure design is to be introduced. This test is called Proper Application Test (PAT). The Steinfurth Closure Tester gives the opportunity to safely perform and observe this test under water in a stable plastic tank.
The second function of the Closure Tester is the Top Load Vent Test (TLVT). This allows testing the resistance against leaking while a strong vertical force is applied to the closure, thus simulating the stacking of the containers (bottles or cans) in trays. This test, too, is performed under water in a tank. The Steinfurth IPPS is used typically as the “pressure engine” for automation of these both tests.
Operation:
PAT: Testing pressure is applied to the specimen closure, which is on the original container (or parts thereof), under water and in the tank. The behavior of the closure (Leakages, popping off or other failures) can be safely observed from outside.
TLVT: A vertical force is applied to the container by a pneumatic cylinder. The behavior of the container (Leakages, cracks, collapsing) can be safely observed from outside.
Product Details | Specification |
Industry | Beverages, Packaging Industry |
Employment | Measuring |
Measurement Categories | Seal security |
Objects | Closures & packages |
Specifics | Test bed |
System | No |
Automation Level | Semi-automatic Machine |
Measuring Methods | Physical Performance Test Method – Secure Seal Test, Physical Performance Test Method – Top Load Vent, Physical Performance Test Method – Vacuum Retention |
- Two important testing methods combined in one device
- Leakages are visible as air bubbles
- Define specimen temperature by heated water
- Stable polycarbonate container
- Supports for glass bottle necks, PET preforms and PET bottles
- Centering weight for bottles
- Safety interlocks
- Pressure range TLVT: 0…10 bar / 0…145PSI
- Pressure range PAT: 0…16 bar / 0…232PSI
- Tank dimensions: 140 mm x 500 mm / 5.5” x 19.7”
- Max. Force (TLVT): 1256 N (approx. 282 lbf)
- Media: Neutral gases