- Home
- Tuotteet
- Savukaasumittaukset
- Durag
- HM 1400 TRX Kokonaiselohopeamittaus
HM 1400 TRX Kokonaiselohopeamittaus
Categories: Tuotteet, Savukaasumittaukset, Elohopeamittaukset, Durag
- General Details
- Overview
- Specification
- Certificate
- Documents
General Details
Total Mercury Monitoring
Continuous, fully-automated analysis of mercury in stack gas emissions (without wet chemicals).
- Cement plants (kilns)
- Coal-fired power generation
- Industrial boilers
- Incinerators (municipal, hospital, sewage sludge wastes)
- Steelworks with scrap metal reconditioning
- Contaminated soil burning plants
- Crematoriums
- Gold and mercury mines
- Fluorescent light bulb recycling plants
Overview
- Automatic calibration
- Mercury speciation (optional)
- Dual range capability (optional)
- M & C Dilution Probe (optional)
- 6-month thermo-catalytic reactor lifetime
- Low cost of ownership
- Simple, safe design
- High reliability
- Robust to chemical interferences
Specification
Measured parameter: | Total mercury |
Measuring Ranges: | 0–45 … 0–500 µg/Nm3 |
Analog Outputs: | 2 (+1*) x 4-20 mA, 500 Ohm |
Digital Outputs: | 4 (+2*) permissable load 48V, 0.5 A |
Digital Inputs: | 1 potential free output |
Detection Limit: | <0.1 µg/dscm |
Calibration Drift: | <3% of measuring range/month |
Zero Point Drift: | automatic zero correction |
Carrier Gas Supply: | 90 psi |
Measuring Principle: | UV-absorption |
Stack Gas Temperature: | 32-480 oF |
Stack Diameter: | No limits |
Supply Voltage: | Monitor: 120/208 V or 120/240 V, 60Hz ¬1, L2, N, PE, 2X 25A, Max 500 VA Heated sample probe/line: 800W Heated sample tube: 800W |
Dimensions (H x W x D): | Cabinet 67 x 32 x 20 in |
Weight: | 485 lbs |
Protection: | IP54 |
Ambient Temperature: | 40 oF to 105 oF (*-10 oF – 120 oF) |
*option |
Certificate
- Suitability-tested by TÜV Hamburg, test report 00 CU 014
- US EPA PS 12
- MCERTS