- Home
- Tuotteet
- Prosessimittaukset
- Teledyne Analytical Instruments
- GFC-7002T / GFC-7002E Ultra Trace Nitrous Oxide (N2O) Analyzer
GFC-7002T / GFC-7002E Ultra Trace Nitrous Oxide (N2O) Analyzer
Teledyne Analytical Instruments
- General Details
- Ominaisuudet
- Specification GFC-7000T
- Specification GFC-7001T
- Specification GFC-7002T
- Documents
The GFC-7002 series gas filter correlation analyzers measure ppb or ppm levels of Nitrous Oxide by comparing infrared (IR) energy absorbed by a sample to that of a reference gas according to the Beer-Lambert law.
Using a gas filter correlation wheel, a high energy IR source is alternately passed through a target gas-filled chamber and a chamber with no target gas present. The light then travels through a variable length folded path to the detector. The energy lost through the sample cell is compared with the reference signal provided by the filter wheel to produce a signal proportional to concentration, with little effect from interfering gases within the sample. This design produces excellent zero and span stability and high signal-to-noise ratio, allowing superior performance over a wide concentration range.
ADVANCED PERFORMANCE
The GFC-7002T offers an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. It includes a built-in data acquisition capability using the analyzer’s own internal memory. This allows the logging of multiple parameters including averaged or instantaneous concentration values, calibration data and operating parameters such as flow, temperature, and pressure.
• Models available for CO, CO2 , and N2 O gas detection
• Gas filter correlation method provides part-perbillion (ppb) detection
• Large color touch-screen provideslarge, luminescent interface
• Adaptive signal filtering optimizesresponse time
• Temperature and pressure compensation
• USB ports on front panel for peripheral devices and firmware upgrades
• Commonality of parts across all models
- 0-200ppb to 0-1,000ppm ranges (configuration dependent), user selectable
- GFC wheel guaranteed against leaks for 5 years
- Continuous monitoring of numerous variables with fully configurable alarms
- Adaptive signal filtering optimizes response time
- Temperature and pressure compensation
- Internal data logging
- Advanced remote operation software
Standard I/O for T-series:
- 4 x analog outputs (up to 3 x 4-20mA current outputs)
- 1 x Ethernet (10 / 100 Base-T)
- 2 x RS-232 (300-115200 baud)
- 2 x USB ports for firmware upgrades
- 8 x opto-isolated digital outputs
- 6 x opto-isolated digital inputs
Optional I/O for T-series:
- 1 x USB communications port
- 1 x RS-485 digital communications port
- 8 x analog inputs (0-10 VDC, 12-bit)
- 4 x digital alarm outputs
- Multidrop RS-232
General | GFC-7000T |
Gas Analyzed | CO2 |
Ranges | 0-2 ppm to 0-2,000 ppm full scale |
(0-100 ppb to 0-100 ppm full scale | |
for –U version) (dual ranges and | |
auto-ranging supported) | |
Measurement Units | ppb, ppm, µg/m3 , mg/m3 (selectable) |
Zero Noise | < 0.1 ppb (RMS) |
(< 2.5 ppb (RMS) with -U option) | |
Span Noise | < 1% of reading (RMS) |
(< 5% of reading (RMS) with -U option) | |
Lower Detectable Limit | 0.2 ppb (< 5 ppb with -U option) |
Zero Drift | < 0.25 ppm/24 hours |
Span Drift | < 0.5% of reading /24 hours |
Lag Time | 10 seconds |
Rise and Fall Time | < 60 seconds to 95% |
Linearity | 1% of fullscale |
Precision | 0.5% of reading |
Sample Flow Rate | 800 cm3 /min ± 10% |
ELECTRICAL SPECIFICATIONS | |
Power Requirements | 100V-120V, 220V-240V, 50/60 Hz |
Analog Output Ranges | 10V, 5V, 1V, 0.1V (selectable) |
Recorder Offset | ±10% |
OUTPUTS | |
Included I/O | 1 x Ethernet: 10/100Base-T |
2 x RS232 (300-115,200 baud) | |
2 x USB device port | |
8 x opto-isolated digital outputs | |
6 x opto-isolated digital inputs | |
4 x analog outputs(3 x 4-20mA current outputs) |
|
Optional I/O | 1 x USB com port |
1 x RS485 | |
8 x analog inputs(0-10V, 12-bit) | |
4 x digital alarm outputs | |
Multidrop RS232 |
General | GFC-7001T |
Gas Analyzed | CO2 |
Ranges | 0-1 ppm to 0-1,000 ppm full scale |
(0-100 ppb to 0-100 ppm full scale | |
for –U version) (dual ranges and | |
auto-ranging supported) | |
Measurement Units | ppb, ppm, µg/m3 , mg/m3 (selectable) |
Zero Noise | < 0.02 ppm (RMS) |
(< 10 ppb (RMS) with -U option) | |
Span Noise | < 0.5% of reading (RMS) above 5 ppm |
(< 0.5% of reading above 2.5 ppm with -U option) |
|
Lower Detectable Limit | 0.04 ppm (< 20 ppb with -U option) |
Zero Drift | < 0.1 ppm/24 hours (< 20 ppb/24 hours with -U option) |
Span Drift | < 0.5% of reading /24 hours (< 0.5% of reading/24 hours above 5 ppm with -U option) |
Lag Time | 10 seconds |
Rise and Fall Time | < 60 seconds to 95% |
Linearity | 1% of fullscale |
Precision | 0.5% of reading |
Sample Flow Rate | 800 cm3 /min ± 10% |
(1800 cm3/min +/- 20% with -U option) /min ± 10% |
|
ELECTRICAL SPECIFICATIONS | |
Power Requirements | 100V-120V, 220V-240V, 50/60 Hz |
Analog Output Ranges | 10V, 5V, 1V, 0.1V (selectable) |
Recorder Offset | ±10% |
OUTPUTS | |
Included I/O | 1 x Ethernet: 10/100Base-T |
2 x RS232 (300-115,200 baud) | |
2 x USB device port | |
8 x opto-isolated digital outputs | |
6 x opto-isolated digital inputs | |
4 x analog outputs(3 x 4-20mA current outputs) |
|
Optional I/O | 1 x USB com port |
1 x RS485 | |
8 x analog inputs(0-10V, 12-bit) | |
4 x digital alarm outputs | |
Multidrop RS232 |
General | GFC-7002T |
Gas Analyzed | CO2 |
Ranges | 0-1 ppm to 0-1,000 ppm full scale |
(0-200 ppb to 0-200 ppm full scale | |
for –U version) (dual ranges and | |
auto-ranging supported) | |
Measurement Units | ppb, ppm, µg/m3 , mg/m3 (selectable) |
Zero Noise | < 0.02 ppm (RMS) |
(< 10 ppb (RMS) with -U option) | |
Span Noise | < 0.5% of reading (RMS) above 5 ppm |
(< 0.5% of reading above 2.5 ppm with -U option) |
|
Lower Detectable Limit | 0.04 ppm (< 10 ppb with -U option) |
Zero Drift | < 0.1 ppm/24 hours (< 25 ppb/24 hours with -U option) |
Span Drift | < 0.5% of reading /24 hours |
Lag Time | 10 seconds |
Rise and Fall Time | < 60 seconds to 95% |
Linearity | 1% of fullscale |
Precision | 0.5% of reading |
Sample Flow Rate | 800 cm3 /min ± 10% |
ELECTRICAL SPECIFICATIONS | |
Power Requirements | 100V-120V, 220V-240V, 50/60 Hz |
Analog Output Ranges | 10V, 5V, 1V, 0.1V (selectable) |
Recorder Offset | ±10% |
OUTPUTS | |
Included I/O | 1 x Ethernet: 10/100Base-T |
2 x RS232 (300-115,200 baud) | |
2 x USB device port | |
8 x opto-isolated digital outputs | |
6 x opto-isolated digital inputs | |
4 x analog outputs(3 x 4-20mA current outputs) |
|
Optional I/O | 1 x USB com port |
1 x RS485 | |
8 x analog inputs(0-10V, 12-bit) | |
4 x digital alarm outputs | |
Multidrop RS232 |