Ametek 5830 Moisture Analyzer

General Details

Model 5830 Moisture Analyzer

 

Building upon the success of our quartz-crystal moisture measurement technology, AMETEK’s Model 5830 is an easy-to-use process moisture analyzer that offers a truly remarkable combination of performance features: exceptional accuracy, multi-gas compatibility, fast response speed, and wide measurement range.

The 5830 is ideal for critical moisture measurement applications including high purity gas production, semiconductor gases, and industrial gas production and quality assurance. Equipped with an on-line verification system, this state-of-the-art analyzer is designed to rapidly build and maintain operator confidence in its analyses. The verification system allows you to challenge the analyzer’s sensor at will with either a zero gas or an actual NIST-traceable moisture concentration to check both its baseline stability and responsiveness to real moisture. No other manufacturer offers these exclusive abilities in their moisture analyzers.

 

Specification

Model 2850 Specifications

Range: Operating range 0.1 to 1000 ppmv. Factory calibrated rante 1 to 100 ppmv with trend indication above and below the calibrated range. (Display will not display values below 0.1 ppmv) Two custom settable 4 to 20 mA outputs are available which may be assigned to cover any portion of the calibrated range, such as 1 to 100 ppmv
Outputs: Four line x 20 character vacuum fluorescent digital display
Two fully progammable 4 to 20 mA analog outputs, into 1200 ohm load
RS485 bidirectional serial port
Alarms: Three pairs: system, alarm, range alert/or calibration alert, concentration alert/or calibration alert.  32 volt, 1 amp noninductive (failsafe by default)
Sensitivity: 20 ppbv or 0.5% of insturment range, whichever is greater
Response Time: 63% to a step change in less than 5 minutes
Accuracy: ±0.05 ppmv or ±5% of the reading whichever is greater
Operating Pressure: Atmospheric
Inlet Pressure: 1 to 10 bars (15 to 150 psig)
Exhaust Pressure: Atmoshpheric
Sample Flow Requirements: less than 600 mL/min at STP
Sample Gas Temperature: 0º to 100ºC (32º to 212ºF)
Ambient Operating Range: 4º to 40ºC (40º top 105ºF), 90% relative humidity, non-condensing, non-corrosive atmoshphere
Software Features: Displays ppmv moisture reading or dew point timer status and instrument status
Power Requirements: 85 to 265 volts, 47 to 63 Hz
Mounting Options: Available for stand-alone or 19″ rack installation
Dimensions (WxHxD): 43.2 x 13.2 x 38.1 cm (17 x 5.2 x 15 in)
Rack-mount version: Same as above except height is 22.1 cm (8.71 in)

 

Certificate
  • NEC Div 2
  • CE