- Home
- Tuotteet
- Prosessimittaukset
- Teledyne Analytical Instruments
- 6400T / 6400E Sulfur Dioxide (SO2) Analyzer
6400T / 6400E Sulfur Dioxide (SO2) Analyzer
Teledyne Analytical Instruments
- General Details
- Options
- Specification 6400T
- Specification 6400T H2S
- Specification 6400TH
- Documents
Teledyne Analytical Instruments (TAI) 6400T uses a proven UV fluorescence principle coupled with state-of-the-art Microprocessor technology to provide accurate and dependable measurements of low level SO2.
The 6400T-H2S measures hydrogen sulfide (H2S) by incorporating a catalytic converter set at 600°F (315°C) to oxidize H2S into SO2 while leaving other more difficult to oxidize sulfurs unconverted. A switching mode allows analysis of both H2S and SO2 concurrently.
Exceptional stability is achieved with the use of an optical shutter to compensate for PMT drift and a reference detector to correct for changes in UV lamp intensity.
A hydrocarbon “kicker” and advanced optical design combine to prevent inaccuracies due to interference. The multi-tasking software gives real time indication of many operational parameters and provides automatic alarms if diagnostic limits are exceeded.
ADVANCED PERFORMANCE
The 6400T offers an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. It includes a built-in data acquisition capability using the analyzer’s own internal memory. This allows the logging of multiple parameters including averaged or instantaneous concentration values, calibration data and operating parameters such as flow, pressure and lamp intensity.
• User-settable ranges, 0-50 ppb to 0-5,000 ppm
• Large, vivid, and durable color graphics display with touch screen interface
• Comprehensive internal data logging with programmable averaging periods
• Front panel USB connectionsfor peripheral devices and firmware upgrades • Adaptive signal filtering optimizesresponse time
• Temperature and pressure compensation
• Internal Zero & Span check (optional)
• Ability to log virtually any operating parameter
• Continuousself checking with alarms
• Selectable voltage
• Internal pump
• Auto ranging and dual ranges
• 47mm diameter particulate filter
• 8 isolated digitalstatus outputs
• Dual bi-directional RS-232
• Specified AC voltage and frequency 100V – 115V or 220V – 240V 50 Hz or 60 Hz
• Up to 4 analog outputs: 4-20mA (3 max) orsettable DC voltage (0-10V, 0-5V, 0-1V, 0-0.1V)
General | 6400T |
Detection | SO2 Analyzer |
Ranges | Min: 0-50 ppb, fullscale; Max: 0-20,000 ppb fullscale; userselectable; dual ranges and auto ranging supported |
Measurement Units | ppb, ppm, μg/m3, mg/m3(selectable) |
Zero Noise | < 0.2 ppb (RMS) |
Span Noise | < 0.5% of reading (RMS)above 50 ppb |
Lower Detectable Limit | 0.4 ppb |
Zero Drift | 7 days |
Span Drift | 7 days |
Lag Time | 20 seconds |
Rise and Fall Time | |
Linearity | 1% of fullscale |
Precision | 0.5% of reading above 50 ppb |
Sample Flow Rate | 650 cm3 / min ±10% |
ELECTRICAL | |
Power Requirements | 100V-120V, 220V-240V, 50/60 Hz |
Recorder Offset | ±10% |
OUTPUTS | |
Included I/O | 1 x Ethernet: 10/100Base-T |
2 x RS232 (300-115,200 baud) | |
2 x USB device port | |
8 x opto-isolated digital outputs | |
6 x opto-isolated digital inputs | |
Up to 4 analog outputs: 4-20mA (3 max) orsettable DC voltage (0-10V, 0-5V, 0-1V, 0-0.1V) | |
Optional I/O | 1 x USB com port |
1 x RS485 | |
8 x analog inputs(0-10V, 12-bit) | |
4 x digital alarm outputs | |
Multidrop RS232 | |
PHYSICAL | |
Operating Temperature Ranges | 5 – 40ºC |
Dimensions (HxWxD) | 7”x 17”x 23.5”(178 x 432 x 597 mm) |
Weight | 35.7 lbs(16.2 kg) |
General | 6400T H2S |
Detection | H2 S and SO2 Analyzer |
Ranges | Min: 0-50 ppb H2 S, 0-50 ppb |
SO2 fullscale | |
Max: 0-10,000 ppb H2 S, |
|
0-20,000 ppb SO2 fullscale user | |
selectable; dual ranges and auto | |
ranging supported | |
Measurement Units | ppb, ppm, μg/m3, mg/m3(selectable) |
Zero Noise | < 0.2 ppb (RMS) |
Span Noise | < 0.5% of reading (RMS)above 50 ppb |
Lower Detectable Limit | 0.4 ppb |
Zero Drift | |
Span Drift | <0.5% of FS / 24 hours |
Lag Time | 20 seconds |
Rise and Fall Time | |
Linearity | 1% of fullscale |
Precision | 0.5% of reading above 50 ppb |
Sample Flow Rate | 650 cm3 / min ±10% |
ELECTRICAL | |
Power Requirements | 100V-120V, 220V-240V, 50/60 Hz |
Recorder Offset | ±10% |
OUTPUTS | |
Included I/O | 1 x Ethernet: 10/100Base-T |
2 x RS232 (300-115,200 baud) | |
2 x USB device port | |
8 x opto-isolated digital outputs | |
6 x opto-isolated digital inputs | |
Up to 4 analog outputs: 4-20mA (3 max) orsettable DC voltage (0-10V, 0-5V, 0-1V, 0-0.1V) | |
Optional I/O | 1 x USB com port |
1 x RS485 | |
8 x analog inputs(0-10V, 12-bit) | |
4 x digital alarm outputs | |
Multidrop RS232 | |
PHYSICAL | |
Operating Temperature Ranges | 5 – 40ºC |
Dimensions (HxWxD) | 7”x 17”x 23.5”(178 x 432 x 597 mm) |
Weight | 41 lbs(18.3 kg) |
General | 6400TH |
Detection | High Range SO2 Analyzer |
Ranges | Min: 0-10 ppm fullscale; |
Max: 0-5,000 ppm fullscale; user | |
selectable; dual ranges and auto | |
ranging supported | |
Measurement Units | ppm, mg/m3 (selectable) |
Zero Noise | < 0.1 ppb (RMS) |
Span Noise | ppm, mg/m3 (selectable) 50 ppb |
Lower Detectable Limit | 0.2 ppb |
Zero Drift | < 1 ppm/24 hours |
Span Drift | < 0.5% of fullscale/24 hours |
Lag Time | 5 seconds |
Rise and Fall Time | < 30 secondsto 95% |
Linearity | 1% of fullscale |
Precision | 0.5% of reading above 10 ppm |
Sample Flow Rate | 700 cm3/min ±10% |
ELECTRICAL | |
Power Requirements | 100V-120V, 220V-240V, 50/60 Hz |
Recorder Offset | ±10% |
OUTPUTS | |
Included I/O | 1 x Ethernet: 10/100Base-T |
2 x RS232 (300-115,200 baud) | |
2 x USB device port | |
8 x opto-isolated digital outputs | |
6 x opto-isolated digital inputs | |
Up to 4 analog outputs: 4-20mA (3 max) orsettable DC voltage (0-10V, 0-5V, 0-1V, 0-0.1V) | |
Optional I/O | 1 x USB com port |
1 x RS485 | |
8 x analog inputs(0-10V, 12-bit) | |
4 x digital alarm outputs | |
Multidrop RS232 | |
PHYSICAL | |
Operating Temperature Ranges | 5 – 40ºC |
Dimensions (HxWxD) | 7”x 17”x 23.5”(178 x 432 x 597 mm) |
Weight | 31 lbs(14 kg) |