6400T / 6400E Sulfur Dioxide (SO2) Analyzer

General Details

Teledyne Analytical Instruments (TAI) 6400T uses a proven UV fluorescence principle coupled with state-of-the-art Microprocessor technology to provide accurate and dependable measurements of low level SO2. 

The 6400T-H2S measures hydrogen sulfide (H2S) by incorporating a catalytic converter set at 600°F (315°C) to oxidize H2S into SO2 while leaving other more difficult to oxidize sulfurs unconverted. A switching mode allows analysis of both H2S and SO2 concurrently. 

Exceptional stability is achieved with the use of an optical shutter to compensate for PMT drift and a reference detector to correct for changes in UV lamp intensity. 

A hydrocarbon “kicker” and advanced optical design combine to prevent inaccuracies due to interference. The multi-tasking software gives real time indication of many operational parameters and provides automatic alarms if diagnostic limits are exceeded. 

ADVANCED PERFORMANCE
The 6400T offers an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. It includes a built-in data acquisition capability using the analyzer’s own internal memory. This allows the logging of multiple parameters including averaged or instantaneous concentration values, calibration data and operating parameters such as flow, pressure and lamp intensity.

Options

• User-settable ranges, 0-50 ppb to 0-5,000 ppm
• Large, vivid, and durable color graphics display with touch screen interface
• Comprehensive internal data logging with programmable averaging periods
• Front panel USB connectionsfor peripheral devices and firmware upgrades • Adaptive signal filtering optimizesresponse time
• Temperature and pressure compensation
• Internal Zero & Span check (optional)
• Ability to log virtually any operating parameter
• Continuousself checking with alarms

• Selectable voltage
• Internal pump
• Auto ranging and dual ranges
• 47mm diameter particulate filter
• 8 isolated digitalstatus outputs
• Dual bi-directional RS-232
• Specified AC voltage and frequency 100V – 115V or 220V – 240V 50 Hz or 60 Hz
• Up to 4 analog outputs: 4-20mA (3 max) orsettable DC voltage (0-10V, 0-5V, 0-1V, 0-0.1V)

Specification 6400T
General 6400T
Detection SO2 Analyzer
Ranges Min: 0-50 ppb, fullscale;
Max: 0-20,000 ppb fullscale;
userselectable; dual ranges and
auto ranging supported
Measurement Units ppb, ppm, μg/m3, mg/m3(selectable)
Zero Noise < 0.2 ppb (RMS)
Span Noise < 0.5% of reading (RMS)above
50 ppb
Lower Detectable Limit 0.4 ppb
Zero Drift 7 days
Span Drift 7 days
Lag Time 20 seconds
Rise and Fall Time
Linearity 1% of fullscale
Precision 0.5% of reading above 50 ppb
Sample Flow Rate 650 cm3 / min ±10%
ELECTRICAL
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Recorder Offset ±10%
OUTPUTS
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device port
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
Up to 4 analog outputs: 4-20mA (3 max) orsettable DC voltage (0-10V, 0-5V, 0-1V, 0-0.1V)
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs(0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
PHYSICAL
Operating Temperature Ranges 5 – 40ºC
Dimensions (HxWxD) 7”x 17”x 23.5”(178 x 432 x 597
mm)
Weight 35.7 lbs(16.2 kg)
Specification 6400T H2S
General 6400T H2S
Detection H2
S and SO2 Analyzer
Ranges Min: 0-50 ppb H2
S, 0-50 ppb
SO2 fullscale
Max: 0-10,000 ppb H2
S,
0-20,000 ppb SO2 fullscale user
selectable; dual ranges and auto
ranging supported
Measurement Units ppb, ppm, μg/m3, mg/m3(selectable)
Zero Noise < 0.2 ppb (RMS)
Span Noise < 0.5% of reading (RMS)above
50 ppb
Lower Detectable Limit 0.4 ppb
Zero Drift
Span Drift <0.5% of FS / 24 hours
Lag Time 20 seconds
Rise and Fall Time
Linearity 1% of fullscale
Precision 0.5% of reading above 50 ppb
Sample Flow Rate 650 cm3 / min ±10%
ELECTRICAL
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Recorder Offset ±10%
OUTPUTS
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device port
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
Up to 4 analog outputs: 4-20mA (3 max) orsettable DC voltage (0-10V, 0-5V, 0-1V, 0-0.1V)
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs(0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
PHYSICAL
Operating Temperature Ranges 5 – 40ºC
Dimensions (HxWxD) 7”x 17”x 23.5”(178 x 432 x 597
mm)
Weight 41 lbs(18.3 kg)
Specification 6400TH
General 6400TH
Detection High Range SO2 Analyzer
Ranges Min: 0-10 ppm fullscale;
Max: 0-5,000 ppm fullscale; user
selectable; dual ranges and auto
ranging supported
Measurement Units ppm, mg/m3 (selectable)
Zero Noise < 0.1 ppb (RMS)
Span Noise ppm, mg/m3 (selectable)
50 ppb
Lower Detectable Limit 0.2 ppb
Zero Drift < 1 ppm/24 hours
Span Drift < 0.5% of fullscale/24 hours
Lag Time 5 seconds
Rise and Fall Time < 30 secondsto 95%
Linearity 1% of fullscale
Precision 0.5% of reading above 10 ppm
Sample Flow Rate 700 cm3/min ±10%
ELECTRICAL
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Recorder Offset ±10%
OUTPUTS
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device port
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
Up to 4 analog outputs: 4-20mA (3 max) orsettable DC voltage (0-10V, 0-5V, 0-1V, 0-0.1V)
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs(0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
PHYSICAL
Operating Temperature Ranges 5 – 40ºC
Dimensions (HxWxD) 7”x 17”x 23.5”(178 x 432 x 597
mm)