4060 Gas Chromatographs

General Details

The 4060 Series of Gas Chromatograph (GC) analyzers combine Teledyne’s field proven detector technologies with gas chromatography column separation capability. This enables reliable and repeatable of a single or multiple compounds in a stream containing interfering components. 

Flame-ionization detector (FID) or thermal conductivity detector (TCD) options enable part-per-billion (ppb), part-per-million (ppm) or percent levels of detection depending on application. The analyzer includes an integral temperature-controlled sample system of 316SS and Teflon construction to provide a suitable sample to the detector. Contact TAI for other available materials.

 

Ominaisuudet

 

  • Field-proven microprocessor-based electronics with menu-driven functionality
  • Easy-to-view 5-digit LED display for readings
  • 2 x 20 vacuum fluorescent display for menu options, settings and status
  • 0-1 VDC and 4-20mA (isolated) analog outputs and gas identification outputs
  • RS-232 duplex digital output
  • Two fully-configurable concentration alarms and one system alarm, all with Form-C relay contacts
  • Built-in diagnostic capability
  • Auto-calibration and remote-calibration ready (integral valves available)

 

Common Detectable Gases 4060/FID | 4060/TC
4060/FID 4060/TC
  Acetaldehyde   Argon (purity and impurity levels)
  Benzene   Carbon Dioxide
  Butane   Carbon Monoxide
  Carbon Dioxide – (measurement achieved with methanizer)   Helium
  Carbon Monoxide – (measurement achieved with methanizer)   Hydrogen
  Ethane   Hydrogen Sulfide
  Ethyl Benzene   Krypton
  Ethylene Nitrogen (purity and impurity levels)
  Hexane   Xenon
  Methane
  Methanol
  MTBE
  Propane
  Propylene
  Toluene
  VOCs
  Xylene
Applications

 

  • Ambient benzene, BTEX detection
  • Beverage-grade carbon dioxide: detection of ppb benzene, acetaldehyde, methanol and VCM in beverage-grade carbon dioxide to ISBT specification
  • Methane / non-methane hydrocarbon monitoring
  • Nitrogen or Argon purity
  • C1-C4 levels in natural gas or syngas production
  • LNG impurity detection
  • MTBE monitoring in gasoline
  • Hydrogen detection in a mixed fuel gas

Chromatography Software – Tracs™
Tracs™ software developed by TAI provides all features and functionality needed to view and analyze chromatography data. When installed on a PC computer, Tracs™ collects digitized data fed from the analyzer with which it can plot chromatograms, perform analytical calculations, log data, and save information for future use. 

Features include:

    • Bi-directional communication of analyzer’s range, alarm and timing settings which can be remotely uploaded and downloaded
    • Acquires real-time raw data for continuous display of chromatograms
    • Chromatograms can be superimposed upon one another to verify repeatability
    • Elution peaks can be displayed within their own respective timing frames or individually manipulated for optimal peak scrutinization
    • Statistical summary including analytical calculation of measurements, distribution of values and deviations
    • Data-logging
    • Saving and retrieving of digitized data files

 

Specification 4060/FID | 4060/TC
4060/FID 4060/TC
Sensing Technology Flame-Ionization Detector (FID) Thermal Conductivity Detector (TCD)
Gas and Ranges Hydrocarbon compounds
Trace or percent level detection
(Contact factory to verify application)
Most gases
Trace or percent level detection
(Contact factory to verify application)
Sensitivity 1% of full-scale 1% of full-scale
Accuracy ± 2% of full-scale
(at constant temperature)
± 2% of full-scale
(at constant temperature)
Repeatability 1% of full-scale through all ranges 1% of full-scale through all ranges
Response Time Application Dependent Application Dependent
Operating Temperature 40-110°F (4-43°C) 40-110°F (4-43°C)
Zero and Span Noise Less than 0.5% of full-scale Less than 0.5% of full-scale
Zero and Span Drift 2% of full-scale per week 2% of full-scale per week
Analog Outputs 0-1 VDC and 4-20 mADC isolated (one
output with Gas ID for 3+ gases)
0-1 VDC and 4-20 mADC isolated (one
output with Gas ID for 3+ gases)
Alarms One system alarm and two concentration
alarms. Form-C relays rated @ 3A 250VAC
resistive
One system alarm and two concentration
alarms. Form-C relays rated @ 3A 250VAC
resistive
Supply Voltage 110 or 220 VAC 50/60 Hz 110 or 220 VAC 50/60 Hz
Maximum Power Consumption 600 VA 600 VA
Maximum Load Impedance of
4-20 mA Output
500 ohms 500 ohms
Sample Flow Rate  1 SCFH (0.5 LPM) standard 1 SCFH (0.5 LPM) standard
Dimensions 19” W x 8.75” H x 22.5” D
(483 x 222 x 572mm)
19” W x 8.75” H x 16.5” D
(483 x 222 x 419mm)